5-Part differential instrument - Open tube

Specifications

Principles

  • Flow Cytometry (FCM), Semi-conductor Laser scatter for WBC differentiation and counting
  • Independent Basophil channel
  • Impedance method for RBC, PLT counting
  • Cyanide-free reagent for Hemoglobin test

Main Features

  • 5-part differentiation, 25+4 parameters, 4 DIFF scatter grams and 1 BASO scatter gram, 2 histograms for RBC and PLT.
  • Tri-angle semi-conductor Laser scatter, advanced flow cytometry
  • Compact, powerful and affordable all-in-one solution
  • Only 20 µl sampling volume
  • Up to 60 samples per hour
  • 3 counting modes: venous blood, capillary blood and pre-diluted
  • Large storage capacity: up to 50,000 samples
  • Built-in thermal printer, also support external printer
  • Built-in barcode scanner, easier data transmission and reagent management

Throughput

Up to 60 samples per hour

Sample Volume

Whole Blood: 20ul, Pre-diluted: 20ul

Test Mode

CBC CBC+DIFF

Data Storage Capacity

Up to 50,000 results including numeric and graphical information

Communication

LAN port supports HL7 protocol

 

Performance

Parameters Precision (CV)
WBC ≤ 2.0% 4.0 – 15.0 x 109/L
RBC ≤1.5% 3.5-6.0 x 1012/L
HGB ≤1.5% 11.0 – 18.0 dg/L
MCV ≤1.0% 70.0 – 120.0 fL
PLT ≤4.0% 150.0 – 500.0 x 109/L
Parameters Linearity range
WBC 40 – 100.0 x 109/L
RBC 0 – 8.00 x 1012/L
HGB 0 – 250 dg/L
PLT 0 – 1000 x 109/L

Parameter

  •  WBC, Lym%, Mon%, Neu%, Bas%, Eos%, Lym#,Mon#, Neu#, Eos#, Bas#, RBC, HGB, HCT, MCV, MCH, MCHC, RDW-CV, RDW-SD, PLT, MPV, PDW, PCT, P-LCR,
    P-LCC, LIC%, LIC#, ALY%, ALY#
  • 4 DIFF scatter grams and 1 BASO scatter gram, 2 histograms